Every micrometer has a manufacturing tolerance on its anvil flatness — typically ±0.3-0.6 μm. After thousands of measurements and any knocks or overtightening, the anvils wear and flatness deteriorates. At 1 μm of anvil convexity, the micrometer produces a systematic error of 1-2 μm depending on the workpiece size — which can blow its ±0.003 mm accuracy specification completely.
How do you verify anvil flatness? There is no ordinary measuring instrument that can measure 1 μm flatness on a flat object. Another micrometer is just as inaccurate as the one you want to verify. A dial indicator doesn’t have sufficient resolution. A coordinate measuring machine can measure on a point basis, but the flatness integral requires interpolation.
The solution has been the same since 1665: OPTICAL INTERFERENCE. Isaac Newton discovered the phenomenon with his rings (Newton’s rings), and Fizeau further developed the method in the 1860s. By using the wavelength of light as the unit of measurement, you achieve precision metrology at the 0.1-0.3 μm level with nothing more than a Ø 30 mm glass disc and a sodium lamp.
The physics of interference in three steps
The optical interference flat is a precision-ground piece of optical-grade glass. Both sides are polished flat to 0.15 μm accuracy. This is how the measurement takes place:
- Place the optical flat on the workpiece’s flat surface. For example, on the micrometer’s anvil. Press down very lightly so the flat sits snugly.
- Illuminate the flat with monochromatic light. The standard choice is a sodium lamp with a wavelength of 589 nm (yellow-orange). Alternatively, white light (which produces colored fringes) can be used, but it is less precise.
- Observe the dark/light LINES (fringes) that appear. The number, shape, and distribution of the fringes tell you everything about the surface flatness.
Each dark-light fringe cycle represents 1 wavelength = 589 nm = 0.589 μm total path difference. Since the light travels DOWN and UP through the gap, the effective flatness change per fringe = 589/2 = 294.5 nm ≈ 0.29 μm.
How fringes arise, optically speaking
The underlying physics is interference between two reflected light streams:
- Stream 1: Light reflected from the UNDERSIDE of the optical flat (the side facing the workpiece).
- Stream 2: Light passes through the flat, reflects off the WORKPIECE SURFACE, and travels back through the flat.
If the two streams differ by a WHOLE NUMBER OF WAVELENGTHS, they are in phase → reinforce each other → LIGHT fringe. If they differ by HALF A WAVELENGTH, they are out of phase → cancel each other → DARK fringe.
The path difference between the streams = 2 × the gap thickness between the flat and the workpiece. So:
- 0 nm gap → in phase → light fringe (contact point).
- 147 nm gap → out of phase → dark fringe.
- 294 nm gap → in phase → next light fringe.
- 441 nm gap → out of phase → next dark fringe.
On workpiece surfaces with varying gaps at different positions (i.e. a non-flat surface), a PATTERN of fringes appears. Compare this with a perfectly flat surface, where ideally there would be no fringes at all (just uniform light).
Reading flatness from fringe count
Standard flatness assessment:
Case 1: No fringes at all. The workpiece is FLAT to within 0.3 μm (less than one fringe). Best case.
Case 2: Parallel straight fringes. The workpiece is PARTIALLY WEDGE-SHAPED. The number of fringes is counted across the diameter. For example, 3 fringes → a gap change of 3 × 0.29 = 0.87 μm across a 30 mm diameter. The surface is flat but tilted by 0.87 μm/30 mm = 29 μm/m.
Case 3: Curved fringes (bowed). The workpiece is DISH-SHAPED (concave or convex). The number of fringes is counted from center to edge.
Case 4: Irregular fringes (bumps, dips). The workpiece has LOCAL flatness anomalies. A classic symptom of wear, corrosion, or impact damage.
Case 5: Newton’s rings (concentric colored fringes). The workpiece has a focal point. A classic symptom of a domed anvil.
Why is a sodium lamp precise?
The standard light source for optical interference metrology is a sodium vapor lamp (Na lamp). Not fluorescent, not LED, not halogen — specifically sodium vapor. The reason:
- Very narrow spectral band: A sodium lamp emits almost exclusively at 588.995 nm and 589.592 nm (the D1 and D2 spectral lines). This monochromaticity produces sharp, clearly defined fringes.
- Coherence length: Sodium light has a relatively long optical coherence length (~1 mm), sufficient for interference at flatness-metrology distances.
- Visible yellow color: The human eye is highly sensitive to yellow-orange light (~589 nm). Fringes are easy to count with the naked eye.
- Historical standard: All metrology textbooks and standards reference the Na D-line. This wavelength is fundamental to international measurement standards.
Alternatively, a cheaper light source can be used: a yellow monochromatic LED (~589 nm) or a filtered white-light lamp. Quality is somewhat lower but still practically usable.
Specifications of this Schut interference flat
- Material: Optical-grade BK7 glass or fluorite glass (depending on the batch supplied).
- Diameter: 30 mm.
- Thickness: 12.00 mm.
- Flatness on both sides: ≤ 0.15 μm (quarter wavelength).
- Parallelism between the sides: ≤ 0.3 μm.
- Optical surface: Ground to λ/4 (quarter wavelength) precision. Anti-reflective coating optional.
- For micrometers with a 0.5 mm spindle pitch: Fits standard micrometers in the 25 mm increment series.
With 0.15 μm flatness on the flat itself, it is as close to perfect as practical metrology allows. Any flatness anomaly you observe on the workpiece comes from the WORKPIECE, not the flat.
Practical workflow for micrometer verification
- Clean the micrometer’s anvil and spindle face with isopropyl alcohol. Fingerprints or dust produce false fringes.
- Open the micrometer and place the optical flat on the anvil. Press down lightly with the spindle so the flat sits firmly.
- Position the sodium lamp so the light falls on the flat at a near-vertical angle.
- Observe through the flat toward the anvil. Count the fringes.
- Note the number, shape, and distribution.
- Assess whether flatness is within accepted spec (typically 1-2 fringes = 0.3-0.6 μm flatness is OK for a workshop micrometer).
- Rotate the flat 45° and repeat. This ensures the fringes are not random artifacts.
- Repeat for the spindle face as well.
- Document the result in the calibration log.
Other applications
Verification of sealing surfaces: Hydraulic cylinders, valves, diaphragms. Flatness under 0.5 μm is often required for effective sealing.
Optics manufacturing: Verification of lens surfaces, filter surfaces.
Semiconductor industry: Wafers, mask plates. Sub-micrometer flatness is critical.
Scientific instrumentation: Spectrometer flat surfaces, laser optics.
Calibration lab reference work: Internal standards for optical metrology.
Educational institutions: Technical universities, mechanical engineering programs. Classic optical metrology instruction.
Precision grinding QC: Verifying grinding machine quality by testing finished work.
Maintenance and handling
- Never touch the polished surface with your fingers. Skin sweat contains chlorides that cause microscopic corrosion.
- Use cotton gloves or optical-grade tissue for handling.
- Extremely drop-sensitive. Optical glass is brittle.
- Clean only with isopropyl alcohol and optical-grade tissue. Never use paper towels or ordinary cloth (causes micro-scratches).
- Store in a protective case with a foam interior.
- Calibration verification annually or after suspected damage. Send to an accredited optical metrology lab.
- Lifespan with proper handling: 30+ years.
Comparison with other flatness measurement methods
Optical interference (this one): DKK 1,324. Precision 0.1-0.3 μm. For flatness on Ø 5-100 mm.
Dial indicator sweep: DKK 500-3,000. Precision 1-5 μm. For larger surfaces, mechanical measurement.
Electronic flatness tester: DKK 10,000-50,000. Precision 0.5-2 μm. For automated QC.
Interferometer (professional): DKK 100,000-500,000. Precision 0.05-0.1 μm. For the laser optics industry.
The optical interference flat is the sweet spot for workshop metrology: extreme precision at a laboratory price.
What you get for your money
Schut plan-parallel optical interference flat, single piece Ø 30 mm × 12 mm thick, flatness on both surfaces ≤ 0.15 μm, parallelism between surfaces ≤ 0.3 μm, optical-grade BK7 glass, dedicated to micrometer verification (compatible with the standard 25 mm increment micrometer series with 0.5 mm spindle pitch), supplied in a protective case. DKK 1,324.
A standard tool for quality control of micrometer anvil flatness, as well as general precision flatness verification of small workpieces. Combined with a sodium lamp (DKK 200-500), it provides a complete precision metrology setup for under DKK 2,000 — compared with professional interferometers costing DKK 100,000+. For everyday toolmaker work, calibration lab work, or optics-industry QC, this instrument is one of the most powerful metrology tools per krone.